Which meter should you rent?
The UF801-P reaches further at both ends and does more. The ProLite is $200 a week cheaper. Two questions settle it: does your pipe fall inside DN25 to DN600, and is the job straightforward.
Start with pipe size
This rules the ProLite out, or leaves it in play.
- Below DN25 or above DN600 — Ultraflux UF801-P. It is the only unit that reaches, covering DN10 to DN1000.
- DN25 to DN600 — either meter works. Move to the next question.

Then ask how demanding the job is
The ProLite is a budget meter built for straightforward water applications. Anything that stretches beyond that is the UF801-P. Take the performance meter if any of these apply:
- Fluid above +80 °C. High-temperature probe kits reach +140 °C. The ProLite stops at 80.
- More than a working day on the pipe. The ProLite runs 12 hours on a battery. The UF801-P logs unattended for up to two months.
- Thermal energy, BTU or kW as the deliverable. That needs the energy kit.
- You want the data in analysis software. The UF801-P connects to a PC with dedicated Ultraflux software. The ProLite hands you a spreadsheet over USB.
- Difficult pipe — lined, heavily attenuating or awkward. There is a probe kit for it.
If none of those apply and the pipe is clean water between DN25 and DN600, the ProLite does the job for $699 instead of $899.
Side by side
| Ultra ProLite | Ultraflux UF801-P | |
|---|---|---|
| Positioning | Budget, straightforward water | Advanced, performance |
| Pipe range | DN25–DN600 (1–24 in) | DN10–DN1000 (0.4–40 in) |
| Weekly rate | $699 | $899 |
| Max fluid temperature | +80 °C | +140 °C with high-temp probes |
| Data readout | Spreadsheet over USB | Dedicated PC software |
| Battery | Up to 12 h continuous | Up to 2 months on sleep |
| Field accuracy | ±1–2% | ±1–2%, ~0.5% field calibrated |
| Reynolds correction | — | Automatic, DSP |
| Thermal energy | If temperatures known | Optional BTU energy kit |
| Difficult / lined pipe | — | Dedicated probe kits |
| Weight | 230 g | Portable field case |
| Wall thickness gauge | 2–24 mm, optional probe | — |
Typical picks
Fire pump annual test, 6″ riser
Ultra ProLite. One day on site, small enough to carry up to the pump room. Fire pump testing
Chilled water balance, 40-coil building
Ultra ProLite. Ordinary water in DN25–DN600 pipe is exactly what the budget meter is for. Chilled water balancing
Water loss audit, 32″ distribution main
UF801-P. Above DN600 it is the only unit that reaches, and it logs for weeks unattended. Water loss auditing
Campus district cooling energy check
UF801-P with the energy kit. Flow and BTU from one instrument. District cooling
Still not sure? Tell us the pipe.
Pipe size, wall material, fluid and what you are trying to prove. We will tell you which unit to take, or that neither will work.